Recommended Separation Distance - Icare PRO TA03-026-1.0 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 34
GUIDANCE AND MANUFACTURER'S DECLARATION-ELECTROMAGNETIC IMMUNITY
Icare PRO (TA03) is intended for use in the electromagnetic enviroment specified below. The customers
or users of Icare PRO (TA03) should assurre that it is used in such enviroment.
Immunity test
Voltage dips, short
interruption and
voltage variations on
power supply lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Radiated RF
IEC 61000-4-3
Conducted RF
IEC 61000-4-6
IEC 60601 Test level Compliance
level
<5 % UT (>95 % dip
NOT
in UT) for 0,5 cycle
APPLICABLE
40 % UT (60 % dip
in UT) for 5 cycles
70 % UT (30 % dip
in UT) for 25 cycles
<5 % UT (>95 % dip
in UT) for 5 s
3 A/m
3 A/m
3 V/m
3V/m
80MHz to 2.5 GHz
3Vrms
NOT
150 kHz to 80 MHz
APPLICABLE
Electromagnetic environment-
Guidance
Icare PRO (TA03) tonometer is not
operational when connected to an
external power supply
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Icare PRO (TA03), including cables,
than the recommended separation
distance calculated from the
equation applicable to the frequency
of the transmitter.
Recommended separation
distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2 5 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey should
be less than the compliance level in
each frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
English
11

Publicidad

Tabla de contenido
loading

Tabla de contenido