PROBLEM: The head or handle is getting hot.
CAUSE: Unit has been used for too many consecutive cycles.
SOLUTION: Place the SmartLite
down for at least 6 minutes.
PROBLEM: Power output indicated on the cradle does not match the results from another power meter.
CAUSE: The cradle power meter is not a quantitative measuring device.
SOLUTION: The power meter incorporated into the charging cradle is meant as a simple diagnostic tool only,
and is calibrated to the output performance of the SmartLite
not give a similar result on the SmartLite
does. Due to the oval shape and dispersion angle of the SmartLite
measured on another power meter may not be accurate.
PROBLEM: The power meter doesn't show a green bar when the output is measured.
CAUSE: SmartLite
®
Max Light head is not placed in the sensor correctly.
SOLUTION: Make sure the SmartLite
against the sensor face.
CAUSE: SmartLite
®
Max Light is in Pulse mode.
SOLUTION: Do not measure the power output in Pulse mode. Select another mode to measure the power output.
PROBLEM: I can't select the time in Boost mode.
CAUSE: Timer function is not available in Boost mode.
SOLUTION: If more time is desired, either use the Boost mode repeatedly at the 5 second curing cycle, or
select another function and choose the desired curing time.
PROBLEM: I press Start, but no light comes out.
CAUSE: Head is not attached properly.
SOLUTION: Make sure the head is pushed completely into the handle, such that the head connector is not
visible.
PROBLEM: I've been using the unit for several cycles and it suddenly stopped working.
CAUSE: Temperature overload – sensor has activated.
SOLUTION: The SmartLite
from excessive heat. If this occurs, place the unit in the cradle or on the counter and wait at least
6 minutes for the unit to cool.
CAUSE: Battery has been drained.
SOLUTION: The battery has been drained of its stored energy. Place the unit in the charging cradle, or attach
the power supply to the handle.
PROBLEM: The red battery low indicator light is on and I plugged in the power cord, but the light still
won't function.
CAUSE: The light has to reset when the battery low indicator light is on continuously and the power cord is
plugged in for corded use.
SOLUTION: plug in the power cord and let the light reset for several minutes before use.
If the above steps do not solve the problem, contact the dealer or point of purchase. DO NOT USE THE
SmartLite
®
Max Light AND DO NOT ATTEMPT TO SERVICE THE UNIT. Serious injury may result.
SmartLite
®
Max L.E.D. Curing Light complies with the following Directives/Standards:
• Medical electrical equipment Part 1: General requirements for safety (IEC 60601-1)
• Standard for Safety for Medical Electrical Equipment, Part 1: General Requirements for Safety (UL 60601-1
and CAN/CSA C22.2 No. 601.1-M90)
• Medical Electrical Equipment - Part 1-4: General Requirements for Safety - Collateral Standard:
Programmable Electrical Medical Systems Consolidated (IEC 60601-1-4)
8
®
Max Light in the charger, or set the unit on a counter, and allow it to cool
®
Max Light head is resting securely in the sensor, with the head resting
®
Max Light is equipped with a temperature overload sensor to protect the LEDs
®
®
Max Light power meter as the SmartLite
Max Light. Other curing lights may
®
®
Max Light, the power
Max Light