4. CARE AND MAINTENANCE
4.1 Hygiene
The SmartLite
®
Max Light is a state-of-the-art electronic device. To prevent unnecessary damage to the unit
and to the power supply, these care and cleaning guidelines must be followed.
• Remove and discard barrier sleeve.
Cross-contamination
• Do not reuse single use products. Dispose of in accordance with local regulations.
• Reprocess reusable products as described below.
• Do not remove curing probe from handpiece until handpiece cleaning and disinfection is completed as
outlined.
4.1.1 Cleaning
• Disconnect the power supply plug from the power outlet and charging base unit prior to
cleaning.
• Clean the unit by gently scrubbing with a disposable towel soaked with hot water and soap or
detergent. Do not allow solution to penetrate the casing.
• Always protect lens of the light probes from abrasive materials. Should the surface become
marred or scratched, light transmission and curing performance will be diminished.
4.1.2 Disinfection
• Probe, handpiece, light shield and/or charging base exposed to spatter or spray of body fluids, or touched
by contaminated hands or oral tissues, should be disinfected with a hospital-level disinfectant according to
national/local regulations. Water-based phenolics and dual or synergized quaternary ammoniums are
approved disinfectants.
• Do not immerse units in disinfectant. Disinfect the units by spraying disinfectant solution onto a paper towel
and wiping or wiping with impregnated wipes. Do not spray disinfectant directly onto the units or allow
solution to penetrate the casing.
• Use of iodophor, chlorine or glutaraldehyde based disinfectant solutions will void the SmartLite
warranty.
• DO NOT immerse unit in water or disinfectant. Wipe – do not spray – solution (with curing probe in place)
onto plastic parts. Prevent liquids from entering openings on unit.
• DO NOT autoclave or otherwise sterilize these components.
• Reconnect the power supply plug to the power outlet and the charging base unit when cleaning and
disinfection process is completed.
4.2 Head replacement
The head portion of the SmartLite
grasp the head firmly and pull it away from the lower handle portion. Attach the replacement head by inserting
the connector into the top of the handle. Make sure you push the head all the way against the handle to ensure
a secure connection. Make sure the power supply is disconnected from the unit before replacing the head.
5. TROUBLESHOOTING
PROBLEM: There is no display.
CAUSE: The unit is in hibernation mode.
SOLUTION: Press the on/off button to wake the unit.
PROBLEM: The unit does not appear to have any power.
CAUSE: The battery is drained.
SOLUTION: Attach the power supply to the SmartLite
the charging cradle, making sure the power supply is attached to the cradle.
CAUSE: The power supply is not attached to the SmartLite
SOLUTION: Make sure the power supply is securely connected to electrical supply outlet and to the
SmartLite
CAUSE: The SmartLite
SOLUTION: Press any button once to bring the unit out of sleep mode. The display should illuminate when this
is done.
®
Max Light may be removed for replacement or cleaning. To do this, simply
®
Max Light handle or cradle.
®
Max Light is in sleep mode.
®
Max Light handle, or place the SmartLite
®
Max Light or the electrical outlet.
®
Max Light
®
Max Light in