8. Error messages
Error that has occurred
'Lo-1' is shown on the display.
'Lo-2' is shown on the display.
'Err' is shown on the display.
„FErr"
„FULL"
No display.
Measurement does not start.
Incorrect weight measurement.
9. Caring for the scale
Only use a damp cloth to clean the scale. If required, you can spray a little cleaner on the cloth. Do not use any aggressive cleaners and never
immerse the scale in water.
Do not clean the scale in the dishwasher!
The electrodes must not be cleaned with alcohol or disinfectant.
10. Warranty conditions
We offer a 3-year warranty for this premium scale from the day of purchase and in accordance with the conditions listed below.
Claims must be made during the warranty period. The date of purchase must be documented by an appropriately completed and stamped warranty
certificate or proof of purchase.
Within the warranty period, HARTMANN will replace or repair free of charge any faulty device components caused by material faults or manufactur-
ing errors. This does not extend the warranty period.
The device is intended for the purpose described in this instruction manual only.
The warranty does not cover damage resulting from improper use or unauthorised interference. Accessory parts that are subject to wear and tear
(e.g. batteries) are excluded from the warranty. Claims for compensation are limited to the value of the merchandise. Compensation for consequential
damage or injury is expressly excluded.
In warranty cases, please send the scale, together with the fully completed and stamped warranty certificate or purchase receipt, directly or via your
dealer to the Customer Services department for your country.
Possible causes
The batteries in the scale (platform) are
empty.
The batteries in the control unit are
empty.
The maximum capacity of 180 kg has
been exceeded.
No contact or only poor contact with
the electrodes.
All memory positions have been
allocated.
The batteries in the control unit are
completely empty.
The batteries are not correctly inserted
in the control unit.
The control unit was not activated
before standing on the scale.
Remote connection between the scale
and the control unit is faulty.
The control unit is too far away from
the scale.
The batteries in the platform are empty
or incorrectly inserted.
Scale is on carpet.
Scale do not have the correct zero
point.
Remedy
Replace the batteries in the platform.
Replace the batteries in the control unit.
The scale can only carry up to 180 kg.
Measure with bare feet or dampen the feet.
Transfer the data to the Veroval
is not transferred, the oldest value will be overwritten.
Change the batteries (see Chapter 4: Settings before the measure-
men).
Check that the polarity is correct (see Chapter 4: Settings before the
measurement).
Activate the scale correctly (wait for '0.0 kg' to appear) and repeat
the measurement.
Switch the control unit on before the measurement.
If necessary, wait 30 seconds and then repeat the measurement.
Re-insert the batteries in the scale and the control unit. Maintain
a greater distance from the following electromagnetic sources:
high-voltage pylons, televisions or computer radio systems, fitness
devices, mobile telephones, weather stations.
Reduce the distance between the scale and the control unit,
maximum guaranteed distance is 2 metres.
Insert full batteries correctly.
When standing on the scale platform, an acoustic signal sounds.
Place the scale on an even and solid surface.
Activate the scale until an acoustic signal sounds. Wait about 4
seconds. Repeat the measurement.
33
English
medi.connect software. If the data
®