7.4 ISM data
SevenGo Duo pro™ meter incorporates Intelligent Sensor Management (ISM
nious functionality provides extra security, safety and eliminates mistakes. The most important features
are:
Extra security!
After connecting the ISM
i al number are transferred from the sensor chip to the meter. The data is also printed on the GLP
print out.
After calibration of the ISM
sen sor chip. The most recent data is always stored where it should be – on the sensor chip!
Extra safety!
After connecting the ISM
be reviewed to see the development of the sensor over time. This information provides an indication if the
sen sor should be cleaned or renewed.
Eliminate mistakes!
After connecting an ISM
Additional features are described below.
Initial calibration data
When an ISM
ferred. The following data is included:
Response time
Temperature tolerance
Membrane resistance
Slope (at pH 7) and offset
Type (and name) of electrode (for example, InLab Expert Pro ISM
Serial number (SN) and ordering (ME) number
Production date
Calibration history
The last 5 calibrations data stored in ISM
ferred.
Maximum temperature
The maximum temperature that the ISM
auto mat i cal ly and can be reviewed for the evaluation of the electrode lifetime.
Reset ISM
®
The calibration history in this menu can be deleted. This menu is protected by a deletion PIN. Upon deliv-
ery, the PIN for deletion is set to 000000. Change the PIN to prevent unauthorized access.
pH/ORP/Ion/Conductivity meter SG78
sensor, the sensor is automatically recognized and the sensor ID and ser-
®
sensor, the calibration data is automatically stored from the meter to the
®
sensor, the five most recent calibrations are transferred to the meter. These can
®
sensor, the last set of calibration data is automatically used for measurements.
®
sensor is connected, the initial calibration data in the sensor can be reviewed or trans-
®
®
sensor has been exposed to during measurement is monitored
®
)
®
sensor including current calibration can be reviewed or trans-
Data management