Performing a Quality Control Test
Performing test with control solution
1. Please run check key test first. After see "
check key from meter. The screen shows the "
need to put a new strip into the meter.
2. Remove test strip from vial and
Re-cap
the vial cap immediately.
3. Insert the test strip with view window, facing up, into test strip port. The code number will appear.
4. The test strip symbol stops blinking. You will see blinking
"
" symbol o
n the screen. Then you need to approach your control solution to the meter.
52
" symbol and "
" appear on the screen. Remove
" symbol blinking and "
" appears. Then you
"
" symbol and
PRECAUTION
- Before you do control solution test, you have to do check key test first.
- Before "
" appears, please don't touch the control solution to the sample entry on strip because
meter is stilling in a internal test. If you do so, the meter will show "
5. Shake the bottle of control solution well before opening the cap. Then open the cap and put it on the table.
6. Drip a drop of control solution on the top of the cap.
NOTE
- Don't touch the control solution by approaching the vial to the sample entry on strip.
- Don't let control solution touch anywhere of your body. In case please use water to clean down.
7. Gently touch sample entry of the strip with the control solution on the top of the cap.
Performing a Quality Control Test
" and "
" .
OPEN
×
53