In the above cases, you must repeat the measurement. Make
sure that the cuff tube is properly inserted and that you do
not move or talk. Re-insert the batteries if necessary, or else
replace them.

9. Specifications

Type:
BM 35/1
Model no.
BM 35
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the upper arm
Measurement range
Cuff pressure 0 – 300 mmHg,
systolic 40 – 280 mmHg,
diastolic 40 – 280 mmHg,
Pulse 40 – 199 beats/minute
Display accuracy
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 135 mm x W 105 mm x H 53 mm
Weight
Approx. 210 g (without batteries)
Cuff size
23 to 36 cm
Permissible operating
+10 °C to +40 °C, 15-90 % relative air
conditions
humidity (non-condensing)
Permissible storage
conditions
Power supply
Battery life
Accessories
Classification
The serial number is located on the device or in the battery
compartment.
Technical information is subject to change without notification
to allow for updates.
• This unit is in line with European Standard EN 60601-1-2
and is subject to particular precautions with regard to elec-
tromagnetic compatibility (EMC). Please note that portable
and mobile HF communication systems may interfere with
this unit. More details can be requested from the stated Cus-
tomer Service address or found at the end of the instructions
for use.
• This device is in line with the EU Medical Devices Directive
93/42/EEC, the „Medizinproduktegesetz" (German Medi-
cal Devices Act) and the standards EN 1060-1 (non-inva-
sive sphygmomanometers, Part 1: General requirements),
EN 1060-3 (non-invasive sphygmomanometers, Part 3: Sup-
21
- 20 °C to +55 °C, 15-95 % relative air
humidity, 700 –1060 hPa ambient
pressure
4 x 1,5 V
AAA batteries
For approx. 250 measurements,
depending on the blood pressure level
and/or pump pressure
Instruction for use, 4 x 1.5 V AAA batter-
ies, storage pouch
Internal supply, IPX0, no AP or APG,
continuous operation, type BF applied
part