USER GUIDELINES

A-1 • This products is designed for dental bone surgery (implant surgery). Please use the product
in a proper use and method.
• Type of protection against electric shock : Class I equipment
• Degree of protection against electric shock : Not classified Type BF applied parts
• Classification according to the degree of protection against ingress of water as detailed in
the current edition of IEC 60529 : Equipment : IPX0, ordinary equipment Foot Switch : IPX8
• This equipment is not suitable for use in the presence of flammable anesthetics or oxygen
• Mode of operation : continuous operation with intermittent loading.
(Loading time : max. 2 minutes / Resting time : min, 15 minutes.)
A-2 • Please pay attention to the use considering the patient's safety first of all.
A-3 • Please read the instruction manual thoroughly for understanding the functions and then use
the product.
A-4 • Check the working status before the surgery and use the product if there is no problem.
A-5 • Stop working and check the problem if there is vibration, heating or strange sound before or
during the use.
A-6 • Ask to the manufacturer if the problem is abnormal or unsolved easily.
A-7 • Please prepare the extra consumable parts before use. Especially, the contra angle handpiece
doesn't work if the blood or a saline solution comes into the head. So that we recommend to
have an extra for safety.
A-8 • Make sure to put a plug in an outlet with earth.
A-9 • Change or insert the bur after the motor stops completely. Otherwise, the inside of contra
angle handpiece could be damaged by the bur and it will cause an accident.
A-10 • Do not drop from on high or give shock on the product. It damages the product.
A-11 • In case the irrigation pump doesn't work, do not bend or fold the tube. The tube will be
broken or taken off.
A-12 • Do not disassemble or modify the product. When disassembled, the after sale service is not available.
A-13 • Please sterilize and lubricate the handpiece immediately after use. Otherwise, the product
could be damaged because of coagulation by the blood or remains from surgery.
A-14 • Do not insert the oil into the motor, It damages the bearing or causes heating.
A-15 • The control unit and foot control pedal are not autoclavable.
A-16 • If the control unit is contaminated, clean with wet cloth and then remove the moisture with
dry cloth after turning the power off.
A-17 • Do not use the solvent such as sinner or benzene for cleaning.
A-18 • The input power is available at AC100~120V / 220~240V. (Check input voltage before
connecting the AC power.)
A-19 • Dispose the irrigation tube after use as medical waste.
It could be a main cause of the breakdown or infection.
When using drill bit, follow the instruction for use provided by the drill manufucturer.
Using it over speed limit may cause an accident.
If display is malfunctioning contact to manufacturer or contact to the place purchaded
Otherwise it may cause an accident.
Use the authenticated product and consumable parts by manufacturer.
Otherwise, it may cause a malfunction or an accideut
OperatingCondition
1) Temperature : 10
2) Humidity : 30% ~ 75%
Be careful not to drop liquid on controlbox.Do not wipe the controller (membrane) with wet fabric.
~40
Typ BF