Internal Calibration | Adjustment
Note:
Only for models with the label Entris..i-1S !
Set the following parameters:
SETUP - BAL.SCAL.- CAL.Just. - Cal.Int. (Code 1.1. 9. 4)
The built-in motorized calibration weight located in the balance housing is applied and
removed automatically for internal calibration.
Step
1. Tare the balance
2. Start calibration
The internal calibration weight
is applied automatically.
3. Adjustment carried out
4. The internal weight is
defined automatically
Key (or instruction) Display
w
Q
CAL.Int.
CAL.RUN.
CAL.end
0.0
g
0.0
g