Battery Charging; Operating Status Display Of The Charger; Power Level Display Of The Handpiece; Operation - 3M ESPE Elipar DeepCure-S Instrucciones De Uso

Lámpara de fotopolimerización dental de tecnología led
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 42

Battery Charging

The unit contains a powerful lithium-ion recharge-
able battery. The battery does not have any
memory effect and can therefore be recharged
at any time by placing it in the charger (see the
section describing the battery power level display
on the handpiece).
The battery can also be charged in the charger
independently of the handpiece.
Place the handpiece on the charger for approx.
1.5 hours before the first use, in order to charge the
new battery completely for the first time.
- The status LED showing the operating status of
the charger flashes green during the charging
operation; see below "Operating Status Display of
the Charger".

Operating Status Display of the Charger

Status LED
Handpiece/battery Handpiece/battery
outside the charger in the charger
Steady green light
Charger ready for

operation

Flashes green
––
Steady red light
Charge contact
pins are wet
Flashes alternately Malfunction in the
red and green
charger

Power Level Display of the Handpiece

Power level LED
Status LED showing
operating status
Operating status
Charging has been
completed
Battery is being
charged
Charge contact
pins are wet
Problem during
charging
Power level LED
Handpiece outside Handpiece in the
the charger
Steady green light
Battery is charged, Not possible,
handpiece ready
for operation
Steady red light
Warning for low
battery capacity,
remaining capacity sleep mode
adequate typically
for 5 10-sec cycles
Flashes red
Battery fully
discharged,
exposure cycle
will be completed
or, if in continuous
mode, stopped
Operation

Selection of Exposure Time

{
LED showing
exposure times
Exposure time options: 5, 10, 15, 20 sec, continuous
mode (120 sec), tack-cure mode.
See the instructions for use for the specific dental
material when selecting the exposure time.
The indicated exposure times assume that the light
guide is held at the exact position over the material
being polymerized.
If the distance between the light guide and the
restoration is increased, the exposure time must be
adjusted accordingly because the light intensity
weakens (see graph).
100
90
80
70
60
50
0
1
2
3
Distance (mm)
Operating status
charger
handpiece is in
sleep mode
Not possible,
handpiece is in
Problem in
charging, battery
is defective or
cannot be charged
START button
TIME button
4
5
6
7
8
9 10
5

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Elipar s10

Tabla de contenido