Tabla De Contenido - 3M ESPE Elipar DeepCure-S Instrucciones De Uso

Lámpara de fotopolimerización dental de tecnología led
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 42
en
ENGLISH
Elipar
DeepCure-S
LED Curing Light
Table of Contents
Safety
from/into the Handpiece
Extended Periods of Non-Use
Safety
PLEASE NOTE!
Prior to installation and start-up of the unit, please read
these instructions carefully!
As with all technical devices, the proper function and
safe operation of this unit depend on the user's
compliance with the standard safety procedures as well
Page
as the specific safety recommendations presented in
1
these Operating Instructions.
2
1. Use of the device is restricted to trained personnel in
accordance with the instructions below. The manu-
2
facturer assumes no liability for any damage resulting
3
from the use of this unit for any other purpose.
3
2. Prior to start-up of the unit make sure that the
3
operating voltage stated on the rating plate is
3
compatible with the available mains voltage. Operation
4
of the unit at a different voltage may damage the unit.
4
3. Position the unit so that the power plug is accessible
4
at all times. The power plug is used to turn the charger
4
on and off. To disconnect the charger from the mains,
4
remove the power plug from the electrical outlet.
4
4. Use only the 3M ESPE charger which is provided with
4
the unit. The use of any other charger can result in
4
damage to the battery.
5
5
5. CAUTION. Do not stare at source. May be harmful
5
to the eyes. Restrict exposure to the area of the
oral cavity in which clinical treatment is intended.
5
Protect patient and user from reflection and intensive
5
scattered light by taking the appropriate precautions,
6
6
e.g., glare shields, goggles, or coverings.
6. CAUTION! As is the case for all high-intensity light-
6
curing devices, the high light intensity is accompanied
6
by heat generation on the exposed surface. This heat
6
can result in irreversible damage if there is longer
7
exposure in the proximity of the pulp or soft tissue.
7
The exposure times given in the manufacturer's
instructions must be observed exactly to avoid any
8
such damage.
8
Uninterrupted exposure times of the same tooth
8
surface in excess of 20 seconds and direct contact
8
9
with oral mucosa or skin must be strictly avoided.
Scientists working in this field are in agreement
9
that the irritation caused by heat generated during
light curing can be minimized by taking two simple
precautions:
10
• Polymerization with external cooling from an air flow
10
• Polymerization at intermittent intervals
10
(e.g., 2 exposures lasting 10 seconds each instead
10
of 1 exposure lasting 20 seconds).
10
7. Elipar DeepCure-S may be operated only with the
10
supplied light guide or original 3M ESPE Elipar
10
DeepCure-S replacement light guide. The light guide
1

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Elipar s10

Tabla de contenido