Press the RUN/STOP button ( ) on the
E
control unit.
When the handpiece is on, the set
piercing speed stops pulsing and the
piercing speed is continuously dis-
played until the handpiece is switched
off.
If the connected handpiece does not
start:
Observe chapter.
E
To switch off the handpiece:
Press the RUN/STOP button again.
E
7

Cleaning and maintenance

NOTICE
Risk of short-circuit and risk of mild elec-
tric shocks
If components of the device are dismantled
and cleaned when voltage is present, there
is a risk of damage to the electronics.
Disconnect the control unit from the
E
power supply before starting all clean-
ing and maintenance work.
38 EN
NOTICE
Damage caused by fluid
If cleansers or disinfectants get inside the
handpiece, the control unit or the power
supply, a short-circuit can occur. The elec-
trical and mechanical components can
also corrode.
Never immerse the handpiece, control
E
unit or power supply in cleansers or
disinfectants.
Never clean the handpiece, control unit
E
or power supply in autoclaves or ultra-
sonic baths.
Pay attention to compatibility of materi-
E
als when selecting cleansers and disin-
fectants (see chapter 7.2 on page 38).
7.1

Routine maintenance

The device has no self-serviceable parts.
As part of the maintenance of the device,
a regular inspection (biennial service) is
required at intervals of 24  months. The
inspection may only be carried out by
authorized local distributors in accor-
dance with the manufacturer's instruc-
tions. All steps required for the inspec-
tion of the device can be found on the in-
spection sheet included in delivery.
Send the device at regular intervals of
E
24  months to your local distributor to
have it inspected in accordance with
the manufacturer's instructions (bien-
nial service).
7.2

Material compatibility

To clean the device, use a 50% aque-
E
ous 1-propanol solution or a mild soap
solution, which is approved in your
country.
To disinfect the device, use one of the
E
following disinfectants:
Manufacturer
Product
Big Spray
Antiseptics
'new'
Bode Chemie Bacillol
Ecolab
Incidin Foam 1 to 2 mins
Schülke &
Mikrozid Liq-
Mayr
uid
Based on the material compatibility, you
can also assess which other cleansers
and disinfectants are suitable.
Exposure time
1 to 5 mins
30 s to 1 min
1 to 2 mins
artyst NO.1