Functional Check; Figure 1- 2: Probe Functional Check Connections - Tektronix P6245 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Functional Check

After installing the probe on the oscilloscope, a functional check
may be performed using the PROBE COMPENSATION connec-
tions on the front panel of the oscilloscope. See Figure 1- - 2.

Figure 1- 2: Probe Functional Check Connections

1. Connect the probe to the oscilloscope.
2. Set the oscilloscope to display the probe's channel.
3. Using a ground lead and a SMT KlipChip, connect the probe
ground to the PROBE COMPENSATION ground connection on
the oscilloscope.
4. Using a standard tip, hold the probe to the SIGNAL terminal on
the oscilloscope.
5. Press AUTOSET (or adjust the oscilloscope) to display the
calibration waveform.
P6245 Instruction Manual
1- 9

Publicidad

Tabla de contenido
loading

Tabla de contenido