en
Re/X700
Reusable Expiratory Filter for 700 Series Ventilators
REF # G-060525-00
Indicates Flow direction. The Re/X700 is designed to be used only in this direction.
Anticipated filter retirement date to be inserted here.
Directions for use
Description
The Re/X700 reusable expiratory filter reduces contaminants delivered to the patient,
and minimizes the possibility of contaminants re-entering the ventilator, where they
can cause reinfection or cross-infection. The filter is designed for repeated autoclaving.
The see-through plastic shell allows visual inspection of the filter.
Specifications
Filter medium: Minimum filter efficiency of 99.97% retention of particles 0.3 μm
nominal at 100 L/min flow.
Resistance to air flow: Less than 2.5 cmH
Size: 3.5 in. diameter (8.9 cm) x 6.70 in. length (17.0 cm)
Internal volume: Approximately 350 millilitres.
Leakage: Less than 0.01 L/min at 3 psi (211 cmH
Construction: Filter housing and end cap material- Polysulfone plastic.
Connectors: ISO 22 mm male conical connectors.
WARNING
1. To ensure that all breathing circuit connections are leak-tight, complete a
breathing circuit leak test according to ventilator operating instructions each
time the Re/X700 filter is installed on a ventilator.
O at 100 L/min flow when new.
2
O) internal pressure.
2
2