Dräger PIR 7000 Instrucciones De Uso página 17

Ocultar thumbs Ver también para PIR 7000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 26
At regular intervals, which are to be determined by the person responsible for the gas detec-
tion system – recommendation, 6 months:
Check the zero-point and the span calibration.
z
Check signal transmission to central unit and alarm activation.
z
A calibration interval of more than six months is possible under the following conditions: after
z
an action time of a maximum of six months, a check should be made of whether an obstruc-
tion can occur in the gas entrance to the measuring cuvette within the given application, e.g.
due to dust, oil, etc. If a restriction to the function caused by these effects is excluded then
the calibration interval can be prolonged – recommendation: maximum 24 months.
Yearly
Inspection by competent personnel. The inspection intervals are to be individually deter-
z
mined with regard to safety regulations, process control conditions and device-related re-
quirements.
Check the measuring cuvette of the gas transmitter and clean if required
To avoid false alarms during inspection, set the analogue output signal to maintenance sig-
z
nal or ensure that the alarm activation at the central device is locked.
Remove the splash guard and, when required, any additional accessories from the gas
z
transmitter.
Examine the air inlet and air outlet openings for contamination and damage.
z
Examine the windows and mirrors, as well as additional accessories for contamination,
z
clean with water or alcohol and dry using cotton wool or a cloth. Do not scratch the mirrors
or windows!
Attach the splash guard and when required any other accessories to the gas transmitter.
z
Reactivate the analogue output signal if it has been set to maintenance signal. Unlock the
z
alarm activation at the central device once more.
Calibration
To operate the gas transmitter, you can either use a magnetic wand (part No. 45 43 428) or a
computer running Dräger CC-Vision GDS software, or a manual controller made by HART
For calibration purposes, test gases can be applied using the calibration adapter PIR 7000
(part No. 68 11 610), the Flowcell PIR 7000 / 7200 (part No. 68 11 490 / 68 11 910), the Flow-
cell PIR 7000 Duct (part No. 68 11 945), the process adapter PIR 7000 (part No. 68 11 915) or
the process cuvette PIR 7000 (part No. 68 11 415).
Observe the installation instructions in the accompanying IFUs of the respective accessories.
Always calibrate zero-point first, then calibrate span. Always use the calibration gas specified
on the gas transmitter for span calibration.
Handling the magnetic wand:
The gas transmitter contains two contact locations marked by » 0 « and
» S « on the housing. To calibrate, place the magnetic wand on the con-
tact points as shown on the following schematic.
After the respective specified timeout values run out, the process is automatically terminated
without saving the values, and the gas transmitter returns to normal operation.
NOTICE
S
0
NOTICE
Zero-Point Calibration Dräger PIR 7000 / Dräger PIR 7200
Action
Status Indicator
Prepare device for zero-point calibration:
Put the magnetic
green/yellow flash
wand on the » 0 «
alternately and
mark and hold it
quickly
there.
Remove the mag-
green and yellow are
netic wand.
on
Initiate the zero-point calibration:
Place the magnetic
wand within 2 sec-
onds on marking
» 0 « for at least
1 second and then
green/yellow flashes
remove it.
alternately
Place on calibration
adapter PIR 7000.
Feed nitrogen or syn-
thetic air with at least
0.5 L/min onto the
sensor.
Make sure that the
measuring cuvette is
completely purged
®
.
with the chosen zero
gas.
Put the magnetic
green and yellow are
wand on the » 0 «
on
mark and hold it
there.
Remove the mag-
green/yellow flash
netic wand.
alternately and slowly
Display of the zero-point deviation:
Wait approx. 1 to 2
The green status
minutes, until the yel-
light flashes a single
low status light goes
rhythm:
out. The flashing
rhythm of the green
The green status
status light indicates
light flashes a dou-
the current zero-point
ble rhythm:
deviation.
The green status
light flashes a triple
rhythm:
Output Signal
Meaning
measuring mode Unlock device for zero-point
calibration.
measuring mode Device awaits start of calibra-
tion.
(Timeout after approx. 5 sec.)
maintenance
The calibration routine is
signal
started.
(Timeout after approx. 4 min.)
maintenance
Confirm that the gas transmit-
signal
ter was purged with the cho-
sen zero gas.
maintenance
Device determines the current
signal
zero-point deviation.
(Timeout after approx. 30 min.)
maintenance
The zero-point deviation is
signal
less than the set "calibration
range limit".
maintenance
Slight zero-point deviation.
signal
maintenance
The zero-point deviation is
signal
larger than ±3 %LEL.
(Timeout after approx. 30 min.)
17

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pir 7200

Tabla de contenido