Immunity Test
IEC 60601 Test Level
Shock waves
±1 kV on line
IEC 61000-4-5
±2 kV on line and mass
Low pressure and brief
5% Ut (95% dip in Ut) for 0,5 cycles
interruptions.
40% Ut (60% dip en Ut) for 5 cycles
IEC 61000-4-11
70% Ut (30% dip en Ut) for 25 cycles
<5% Ut (>95% dip en Ut) during 5 seconds
Magnetic field to network
3 A/m
frequency (50/60 Hz)
IEC 61000-4-8
RF conducted IEC 61000-4-6
3 Vrms
150 kHz a 80 MHz
RF irradiate IEC 61000-4-3
3 v/m
80 MHz a 2,5 GHz
NOTA 1: At 80 MHz and at 800 MHz, the highest frequency range is applicable.
NOTA 2: These guidelines are not necessary applicable in all situations. The electromagnetic propagation is affected by the absorption and reflections of structures, objects and people.
a The intensity of the fixed transmitters, i.e, the base stations for radiophones (cellular/Wireless) and mobile radios, amateur radio, AM and FM radio, TV, cannot be theoretically foreseen
with precision. To evaluate the electromagnetic environment due to fixed transmitters, it is possible to consider an electromagnetic investigation on site. If the field's intensity, measured at
the location where eKuore Pro® / eKuore Pro® amplified is used, exceeds the level of conformity for RF below.
It is advised to verify that eKuore Pro® / eKuore Pro® amplified is working properly. If abnormal performance is observed, supplementary measures could be necessary, in order to re-
orientate or reposition eKuore Pro® / eKuore Pro® amplified.
b On the frequency range 150 kHz to 80 MHz, it is advised that the field's intensity is less than 3 V/m
Conformity
Electromagnetic environment – Indications
level
Not applicable
The quality of the power should be comparable to that a commercial
Not applicable
establishment or hospital
Not applicable
The quality of the power should be comparable to that a commercial
Not applicable
establishment or hospital.
Not applicable
Not applicable
3 A/m
The magnetic fields to network frequencies should be of the same
level of characteristics to that of a commercial establishment or
hospital
3 Vrms
Portable radio communication must not be used in the vicinity of
150 kHz a 80
eKuore Pro® Series, including cables, unless it is of the recommended
MHz
calculated distance applicable to the transmitter's frequency.
Separation distance recommended
3 v/m
d = [3,5/V1] √P
80 MHz a 2,5
d = [3,5/E1] √P 80 MHz a 800 MHz
GHz
d = [7/E1] √P 800 MHz a 2,5 GHz
Where P is the transmitter's maximum power in Watts (W),
according to the transmitter's manufacturer, and is recommended
separation distance in meters(m). It is appropriate that the fields of
intensity are fixed RF transmitters, determined by an electromagne-
tic investigation in sitea, are lower than the levels of conformity, in
every frequency rangeb. Interferences may take place in proximity
to devices equipped with the following symbol: