Figure 5
Steps
Start Time
Intermittent
Mode
Loop
Click RESET to erase all method values.
5 steps can be set for each method.
All steps are programmable with speed, timer, ramp, and intermittent mode.
A method is considered set when at least speed and timer are set for one step.
Rpm: set point speed
Time: countdown performed for each step visualized also on the main screen
Ramp: if
is selected, OHS stirs for all the time at the speed set
If
is selected, OHS reaches the speed set in time selected
Interm. Mode: if YES is selected, but no intermittent mode is set for the method, OHS works in continuous
mode.
It allows to set an hour at which the method starts.
Enter in the menu, rotate speed control knob (C) till the desired hour. Click the knob to confirm.
Even if a Start Time is set, the method begins if START button in method list is clicked.
It allows to set stirring period and stop period alternatively.
Enter in the menu, rotate speed control knob (C) till the desired working time. Click to confirm.
Rotate speed control knob (C) till the desired pause time. Click to confirm.
Click again the knob to modify set values.
It allows to repeat the whole method N times.
Enter in the menu, rotate speed control knob (C) till the desired value. Click to confirm.
13