PIN / c able assignment VP 8 cable:
VP pin VP8 cable
A
B
C
D
E
F
G
H
S
1 T he 4 / 20 mA output is an active source and
needs to be connected to a passive receiver
e.g., passive PLCard.
2.3
4 / 20 mA output configuration
The configuration of the 4 / 20 mA outputs can be
conducted with iSense PC software or with the
iSense Mobile app (download from Google Play
or Apple App Store).
2.4
MODBUS RTU
For support of the MODBUS RTU programming
please download the MODBUS Programmers
Guide from www.mt.com/pro-M100SM-docu-
mentation.
M100 SM Series
30 370 187 C
M100 SM Sensor Mount Transmitters
black/transparent out1 4 to 20 mA+
red
gray
blue
white
green
pink
brown
green/yellow
Function
out1 4 to 20 mA–
24 VDC+
24 VDC–
out2 4 to 20 mA+
out2 4 to 20 mA–
RS 4 85+ (MODBUS RTU)
RS 4 85– (MODBUS RTU)
shield
© 03 / 18 Mettler-Toledo GmbH
Printed in Switzerland
1
1