Troubleshooting

An error message is displayed in the event of instrument faults. The
number that appears below the message is used for service functions.
If the fault remains after switching on and off several times, contact
DrägerService.
Fault / message
Instrument cannot
be switched on.
Volume too low /
sample invalid.
Instrument outside
temperature range
↑ ↑ ↑
(Measuring range
exceeded)
Sample interruption Irregular outward
Error number
Cause
Batteries are empty
Test subject is blowing
intermittently or not
hard enough.
Instrument too hot or
too cold.
Measuring range
exceeded. Alcohol
residues that are still in
the mouth may cause
incorrect measuring
values.
breath, abrupt end or
intake
Instrument error
Troubleshooting
Remedy
Check batteries
Blow more strongly and
without interruptions
into the funnel.
Allow the instrument to
cool down or get
warmer.
There must be an
interval of at least
10 minutes after alcohol
has been taken into the
mouth.
To repeat, press OK
Repeat breath sample,
by pressing OK.
Switch the instrument
off and on. If the error
remains, contact
DrägerService or your
dealer.
en
18

Publicidad

Tabla de contenido

Solución de problemas

loading