Troubleshooting

5
Troubleshooting
In the event of an instrument malfunction, an error message is displayed. The
number which appears underneath the message is used for service functions. If
the malfunction does not disappear after switching the instrument on and off
several times, then please contact DrägerService.
Malfunction/Error
Instrument will not
switch on.
Volume too small /
sample invalid.
Instrument outside the
temperature range
↑ ↑ ↑
(Measuring range
exceeded)
Sampling interruption
Error number
28
Cause
The batteries are empty.
The test person is blowing
unevenly or not strongly
enough.
The instrument is too
warm or too cold.
Measuring range
exceeded. Alcohol residue
in the mouth can result in
faulty measured values.
Irregular breathing out,
abrupt ending or breathing
in
Instrument error
Remedy
Change batteries.
Blow with more force and
without interruption into
the funnel.
Allow instrument to cool
down or warm up.
At least 15 minutes wait
time must have passed
since alcohol was last
consumed through the
mouth.
To repeat, press the
button
Repeat breath sample by
pressing the
Switch the instrument off
and on again. If the error
remains, contact
DrägerService or the
distributor.
button.