Trouble-Shooting; Environmental Protection; Warranty - Mettler Toledo 3201-UPW Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 18
pH Combination Electrode Type 3201-UPW / 3201 i-UPW
of buffer solution pH 9 to 10 parts of electrolyte, this
in order to avoid corrosion of housing parts.
4. If a sensor is stored dry for a few days by mistake, let
it soak in the normal storage electrolyte for several
hours before use.
5. Occasionally check the connector for possible traces
of moisture. If necessary, clean well with deionized
water or alcohol, and afterwards dry carefully.
7

Trouble-shooting

Sluggish response time (e.g. after a long and dry
storage period) is an indication that reactivation of the pH
sensitive glass may be necessary. Dip the pH sensitive
glass (but no further sensor parts!) into Reconditioning
(HF) Solution (METTLER TOLEDO product, article no. 51
319 001) for 5 min. Afterwards, rinse well with deionized
water and let sensor tip and diaphragm soak in reference
electrolyte for minimum 12 hours.
8

Environmental protection

Waste electrical products should not be
disposed of with household waste. Please
recycle where facilities exist. Check with
your Local Authority or retailer for recycling
advice.
9

Warranty

In the event of manufacturing faults, 12 months warranty
from date of production is granted.
© 05 / 18 Mettler-Toledo GmbH
Printed in Switzerland
3201-UPW / 3201 i-UPW
52 002 471 F
en
17

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

3201i-upw

Tabla de contenido