5.4.3 Carbon Monoxide

Range
Resolution
Reproducibility
Response Time
(Typical)
TEST GAS APPLIED
Hydrogen Sulfide (H 2 S)
Carbon Monoxide (CO)
Nitric Oxide (NO)
Nitrogen Dioxide (NO 2 )
Sulfur Dioxide (SO 2 )
Chlorine (Cl 2 )
Hydrogen Cyanide (HCN)
Ammonia (NH 3 )
Toluene
isopropanol
Hydrogen (H 2 )
5-4
0 - 1999 ppm CO
1 ppm CO for 6 - 1999 ppm H 2 S
± 5 ppm CO or 10 % of reading, whichever is greater
(normal temperature range)
±10 ppm CO or 20 % of reading, whichever is greater
(extended temperature range)
90% of final reading
<15 seconds (normal temperature range)
CONCENTRATION
(PPM) APPLIED
40
100
50
11
9
10
30
25
53
100
100
CO CHANNEL
% CROSS-SENSITIVITY
0
100
84
0
-4
0
-5
0
0
-8
48