Battery And Fuse Replacement - Amprobe AM-540 Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

BATTERY AND FUSE REPLACEMENT

X�
WARNING
To avoid shock, injury, or damage to the Meter:
Disconnect test leads before opening case.
Use ONLY fuses with the amperage, interrupt, voltage, and speed ratings
specified.
Replacing BATTERY follow below steps:
1. Disconnect the test lead probe from measuring circuit.
2. Turn the Meter to OFF position.
3. Remove the screws from the battery cover and open the battery cover
4. Remove the batteries and replace with one 9V (6F22) or equivalent. The
battery cover provides the correct polarity fitting construction design.
Install the battery in the battery cover.
5. Put the battery cover back and re-fasten the screw.
Battery: 9V (6F22) Battery or equivalent
Replacing FUSE follow below steps:
1. Disconnect the test lead probe from measuring circuit.
2. Turn the Meter to OFF position.
3. Remove the screws from the enclosure and open the enclosure.
4. Remove the broken fuse and replace with new specified fuse.
5. Put the enclosure back and re-fasten the screw.
Fuse ratings:
mA /µA input terminal: F1 fuse, 0.5A H 1000V fast-fuse, ( 6×32)mm
10 A input terminal: F2 fuse, 11A H 1000V fast-fuse, ( 10×38)mm
24

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Am-550Am-540-eurAm-550-eur

Tabla de contenido