Battery And Fuse Replacement; Specifications; General Specifications - Amprobe 34XR-A Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 92

Battery and Fuse Replacement

� �Warning
To avoid electrical shock remove the test leads from both the meter and
the test circuit before accessing the battery or the fuses.
To access the battery and the mA fuse remove the two screws holding the
Battery/Fuse Cover in place, and lift the cover from the meter.
To replace the mA fuse, pry it from its clips using a small screwdriver. A spare
mA fuse is located between the battery and the mA fuse.
mA Fuse: Fast Blow 315mA/1000V, minimum interrupt rating 30 kA
(6.3 x 32 mm) (Amprobe
To replace the 10 A fuse: 1) Remove the battery. 2) Remove the four rear-case
screws. 3) Separate the case. 4) Remove the 10 A fuse cover. 5) Remove and
replace the 10A fuse. 6) Re-install the fuse cover. 7) Reassemble the meter.
10A Fuse: Fast Blow 10A/1000V, minimum interrupt rating 30 kA
(10 x 38 mm) (Amprobe® FP100).

SPECIFICATIONS

General Specifications

Display
Polarity
Overrange
Zero
Low battery indication
Auto power off
Measurement rate
Operating environment
Storage temperature
Temperature Coefficient
See Figure -13-
FP300)
®
3 ¾ digit liquid crystal display
(LCD)(3999 count) with a 41-segment
Automatic, positive implied, negative
polarity indication.
(OL) or (-OL) is displayed.
Automatic.
The  is displayed when the battery
voltage drops below the operating level.
Approx. 30 minutes.
2 times per second, nominal.
0 °C to 50 °C at <70 % R.H.
-20 °C to 60 °C, 0 to 80 % R.H. with battery
removed from meter.
0.1 × (specified accuracy) per °C. (0 °C to 18
°C, 28 °C to 50 °C).