Electromagnetic Compatibility Information
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration – electromagnetic emission
The BC 57 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BC 57 should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration – electromagnetic immunity
The BC 57 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 57 should
assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Compliance
Electromagnetic environment – guidance
Group 2
The BC 57 must emit electromagnetic energy in order to perform its inten-
ded function. Nearby electronic equipment may be affected.
Class B
The BC 57 is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
Not applicable
supplies buildings used for domestic purposes.
Not applicable
IEC 60601 test
Compliance
level
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30 %.
Power frequency magnetic fields should be at levels characteri-
stic of a typical location in a typical commercial or hospital envi-
ronment.
138