During operation

During operation, the measured values for every measured gas are displayed.
In the event of an alarm, the corresponding displays, including the visual,
audible and vibration alarms, are activated. See section "Identifying alarms".
If a measuring range is exceeded or not reached, the following displays are
shown instead of the measured value display:
»
«
(measuring range exceeded) or
»
«
(measuring range not reached).
After the measuring range of the TOX measuring channels has been
exceeded temporarily (up to one hour), checking the measuring channels
is not necessary.
Special states in which there is no measuring operation (quick menu,
calibration menu, warm-up of sensors, password input) are indicated by a
visual signal (slow flashing of the alarm LED
If an IR sensor is used in the Dräger X-am 5600, a calibration of zero point and
sensitivity must be carried out after experiencing an impact load that results in
a non-zero display when exposed to fresh air.

Identifying alarms

An alarm is displayed visually, audibly and through vibration in a specific pattern.
At low temperatures the legibility of the display can be improved by switching
on the backlight.
Concentration pre-alarm A1
Intermittent alarm:
Display » A1 « and measured value alternately.
Not for O
!
2
The pre-alarm A1 is not latching and stops when the concentration has dropped
below the alarm threshold A1.
In case of A1, a single tone is audible and the alarm LED flashes.
In case of A2, a double tone is audible and the alarm LED flashes twice.
NOTICE
).
WARNING
NOTICE
Acknowledge pre-alarm:
Press the [ OK ] key, only the audible alarm and the vibration alarm are
switched off.
Concentration main alarm A2
Intermittent alarm:
Alternating display of » A2 « and measured value.
For O
: A1 = lack of oxygen,
2
A2 = excess oxygen.
Risk of fatal injury! Leave the area immediately.
A main alarm is self-retaining and cannot be acknowledged or cancelled.
After leaving the area, when the concentration has dropped below the alarm
threshold:
Press the [ OK ] key, the alarm messages are switched off.
The measuring range 0 to 100 vol. % CH
explosive mixtures in the measuring range of 0 to 100 %LEL.
STEL / TWA exposure alarm
Intermittent alarm:
Display » A2 « and »
« (STEL) or »
Health hazard! Leave the area immediately.
After this alarm, the deployment of personnel is subject to the relevant national
regulations.
The STEL alarm can be triggered with a maximum delay of one minute.
The STEL and TWA alarm cannot be acknowledged or cancelled.
Switch off the instrument. The values for the exposure evaluation are
deleted after the instrument is switched on again.
WARNING
WARNING
is not suitable for monitoring
4
« (TWA) and measured value alternately:
WARNING
NOTICE
7