A P P L I C A T I O N
▷ Using self-curing synthetic material [A],
"detachably" connect the tracing tablet [1] with
E N
the tray. The synthetic material must have com-
plete contact with the lateral notches of the tab-
lets, yet may not encase the notch from above,
as otherwise the tablet cannot be detached.
A
Fig. 4
Mandibular tray with tracing tablet
▷ Affix alignment clamp [6] with locking screw [7]
to tracing tablet [1].
6
Fig. 5
Affixing the alignment clamp
20
7
1
9
To mount the tracing ball on the individual func-
tional tray of the maxilla:
▷ Select long or short tracing ball depending on
space vertically available (here: Tracing ball,
long [4]).
▷ Screw tracing ball together with lock nut [4a]
about halfway into tracer mounting.
This leaves enough threads in both rotation
directions for subsequent adjustment.
▷ Protect the threads of the tracing ball facing
towards the palatinal side with wax or a piece of
the plastic hose [12].
▷ Screw in retention screws [3a] of tracer mount-
ing [3] to the stop. Insulate the threads with
vaseline (petroleum jelly).
4
4a
6a
Fig. 6
Schematic diagram of lock
▷ Position tracing ball [4] (with tracer mounting)
in alignment clamp [6] maximal 6 mm frontal to
chewing center and vertical to the tracing plate,
and then lock it with the screw to the alignment
clamp [6a].
▷ Apply autopolymerizate to the maxillary spoon
in the chewing center area, so that is runs fron-
tal.
▷ Close the articulator.
The tracer mounting is embedded in the autopo-
lymerizate of the maxillary spoon.
3
3a
7
6