5.5. Connection to the facility's water
network or stand-alone heating|cooling
system (applicable to Palletank
®

for LevMixer

)
A. Connect the inlet and outlet of the
Palletank
®
heat transfer fluid circuit to the
heating|cooling system. The inlet and outlet
of the heat transfer fluid circuit are located
on the back of the Palletank
®
B. Open the two manual ball valves and start
filling the heat transfer fluid into the dimple
jacket circuit according to your internal pro-
cedure. The dimple jacket circuit is designed
in such way that the air escape naturally
from the dimple jacket.
Figure 29. Inlet (right) and outlet (left) of
the dimple jacket
Note
The heat exchange system inlet and outlet
comes configured with a Triclamp DN20
(1 ½") connection.
5.6. Coupling the drive unit with Palletank
for LevMixer
®

and Bag

Notice:
Depending on your process requirements,
the LevMixer
®
drive unit can be either cou-
®
pled to the Palletank
for LevMixer
Palletank
®

Jacketed for LevMixer

prior to the filling or once the bag is filled
to its nominal volume:
– If mixing is required along with bag filling:
®
the LevMixer
drive unit will be inserted
prior to filling by following the instruc-
tions below.
Caution:
DO NOT attempt to mix in dry bags. It may
result in damage of the film. The bag must
be filled to 30% of its nominal volume
prior to starting the mixing – Please refer
to Section 6 for mixing instructions.
– If mixing is not required during the filling:
the magnetic clamp must be maintained in
position and the bag will first be filled to
its nominal volume (Refer to section 5.6).
The LevMixer
®
drive unit will be coupled
after all by following the instructions
below.
A. Make sure that the LevMixer
(two or three latches) and Palletank
®
Jacketed
LevMixer
®
A.1. LevMixer
works with Palletank
200L and 400L (See Fig. 30)
(see Figure 29).
Figure 30.
A.2. LevMixer
works with Palletank
200L, 400L, 650L and 1000L. (See Fig. 31)
Figure 31.
B. Remove magnetic clamp from the bag-
tank assembly. To remove the magnetic
®
clamp, reach underneath the drive-bag inter-
face and carefully pull the magnetic clamp
until it is free from the bag-tank assembly
(See Fig. 32). Return the magnetic clamp to
the supplied Accessories Box for future use.
®
or
®
and bag
Figure 32.
®
drive unit
®
for
match:
®
drive unit with two latches
®
for Mixing 50L, 100L,
®
drive unit with three latches
®
for Mixing 50L, 100L,
C. Shift the latch toward the control box
slightly (See Fig. 33).
Figure 33.
D. Carefully press down on the drive handle
and raise the front wheels off the ground.
Align the guide bearings of the LevMixer
drive unit with the guide rails of the
Palletank
®
LevMixer
®
(See Fig. 34).
Figure 34.
E. Roll the bearings of the LevMixer
unit along the rails all the way until the dead
end of the rails (See Fig. 35).
Figure 35.
G. Using the drive unit handle, raise the
®
LevMixer
drive unit to an upright position.
While holding the LevMixer
®
this position shift the smaller latch toward
the Palletank
®
for LevMixer
®
or Palletank
Jacketed for LevMixer
®
so that the cross
bar rests on the grooves in the guide rails
(See Fig. 36).
®
®
drive
drive unit in
®
13