Current Measurement - Greenlee CMI-2000 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Typical Measurements and Accuracy (cont'd)

Current Measurement

AC Current
Range
Accuracy Range
200.0 A
0.0 A to 200.0 A
0 A to 500 A
2000 A
500 A to 1000 A
1000 A to 2000 A
DC Current
Range
Accuracy Range
200.0 A
2000 A
AC + DC Current
Range
Accuracy Range
200.0 A
0.0 A to 200.0 A
0 A to 1000 A
2000 A
1000 A to 2000 A
† Induced error for adjacent current carrying conductors is < 0.1 A/A.
Note: Use Δ if there is an offset before the measurement is taken.
± (2.5% + 0.5 A)
± (3.0% + 5 A)
± (3.5% + 5 A)
unspecified
0.0 A to 200.0 A
± (2.0% + 0.5 A)
0 A to 500 A
± (2.0% + 5 A)
500 A to 2000 A
± (3.0% + 5 A)
± (3.5% + 0.8 A)
± (3.5% + 8 A)
unspecified
Accuracy at
40 Hz to 50 Hz†
Accuracy†
Accuracy at
40 Hz to 50 Hz†
Accuracy at
50 Hz to 60 Hz†
± (2.0% + 0.5 A)
± (2.5% + 5 A)
± (3.0% + 5 A)
± (3.0% + 5 A)
Accuracy at
50 Hz to 60 Hz†
± (3.0% + 0.8 A)
± (3.0% + 8 A)
± (3.0% + 5 A)
CMI-2000
Accuracy at
60 Hz to 400 Hz†
± (2.5% + 0.5 A)
± (3.0% + 5 A)
± (3.5% + 5 A)
unspecified
Accuracy at
60 Hz to 400 Hz†
± (3.5% + 0.8 A)
± (3.5% + 8 A)
unspecified
13