Assembly Of Internal And External Irrigation System (Optional) - Nouvag MD11 Instrucciones De Uso

Ocultar thumbs Ver también para MD11:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

MD 11
6.5

Assembly of internal and external irrigation system (optional)

A
B
Art. No.: 1777
C
Art. No.: 1873
D
D
E
F
G
MD 11 Operating Instructions, Art. No. 31665, V10/14
Art. No.: 1712
C
Art. No.: 1881
A) Graft a 6 cm tube piece (optional, Art. No.: 1773) on
each of the branching tubes of the Y-Connector
(optional, Art. No.: 1777).
C
B) Connect one of the 6 cm tube pieces that branch out
of the Y-Connector with the internal cooling tube
(delivered with the contra angle, or optional, Art. No.:
1712).
C) Secure small, grey clips and the bigger, white clips to
the tube set.
D) Secure bigger, white clips to the contra angle and the
small, grey clips to the motor cable.
E) Put the internal cooling tube with attached tube piece
into the hole in the head of the contra angle and the
drill bit (special drill bits are necessary with a hole for
internal cooling).
F) Push the spigot of the cooling tube into the hole in
the shank of the clamping mechanism.
G) Connect the 6 cm tube piece (optional, Art. No.: 1773)
with the tube of the external cooling.
If needed, secure further clips to the cable of the
motor or the contra angle.
EN
9

Publicidad

Tabla de contenido
loading

Tabla de contenido