Warnings / Safety instructions
lII
The Basic 30 is approved exclusively for the use as described in these instructions. Medela
can only guarantee the safe functioning of the system when the Basic 30 is used in combinati-
on with the original Medela accessories (collection system, tubes, filter etc.).
The Basic 30 is EMC-tested in conformity with the requirements of IEC 60601-1-2:2007 and
can be used in the vicinity of other EMC-tested devices that fulfill the requirements as outlined
in the IEC 60601-1-2 standard. Untested HF sources, radio networks or the like can imapir the
function of the device and should not be operated in combination with the Basic 30.
Please read and observe these warnings and safety instructions before putting into
operation.
Warnings
• For use only by medically trained persons who have been adequately trained in suction
procedures and in the use of aspirators. Caution: incorrect use can cause pain and injury
to the patient.
• Before you plug in the device, please check that your local power supply is the same as
the voltage given on the specification plate.
• The device must not be used for suctioning explosive, easily flammable or corrosive liquids
• The connecting tubing supplied with the device must never come into direct contact with
the suction area.
• Before cleaning the device, pull the plug out of the mains socket.
Safety instructions
• The Medela Basic 30 is a medical device that requires special safety measures in regard
to EMC. It must be installed and put into operation in accordance with the attached EMC
information.
• Portable and mobile RF communications equipment can affect medical devices.
• In the case of overflow, inform the internal technical service immediately and perform the
tasks in the service manual.
• In each of the following cases, the device must not be used and it must be repaired by
Customer Services:
- if the power cord or the plug are damaged
- if the device is not functioning perfectly
- if the device is damaged
- if the device shows clear safety defects.
• Keep the power cord away from hot surfaces.
• The plug and the ON / OFF switch must not come into contact with moisture.
• Separation from the mains is only assured through the disconnection of the plug and socket
connection.
• Never pull the plug out of the mains socket by pulling on the power cord!
• Never leave the device unattended when it is switched on.
These instructions for use must be kept for later reference.
EN
3