Safety Tests
B
B
Safety Tests
DIN EN 62353: 2008-8
Hospital
Location
Visual checks
Cracks, brittle areas, discoloration on housing and display
Basic 30 is clean with no dirty areas
Measurements*
Max Vacuum (use Basic 30 with 1 L jar and tubing set)
Set vacuum to - 85 kPa (-638 mmHg).
Using a short piece of suitable tubing (7 mm ID) connect
suction tube to vacuum gauge and start unit, wait 1 minute for
the pump to reach equilibrium
Max. Flow (keep settings from above)
Disconnect vacuum gauge and connect flow meter instead,
wait 2 minutes for the pump to reach equilibrium
Overall evaluation
No safety or fuctional technical faults were detected.
Faults that are not safety relevant were detected. No direct risk.
Safety relevant faults were detected.
* Required Test Equipment
Vacuum gauge Class 2.5 % or better (074.0006)
Flow Meter air 20°, 1 bar abs, Class 4 % or better
Tester / Signature
146
Basic 30-Serial number
OK
❑ Measured value is
between
-85 kPa
(-638 mmHg)
+/- 5%
❑ ~ 30 L/min
Faults
❑
❑ 1 ________________
2 ________________
3 ________________
❑ 1 ________________
2 ________________
3 ________________
Next test
Place / Date
NO
YES
Not OK
❑ Measured value
_________________
❑ Measured value
_________________
Measures
Basic 30 is ready
for use
Basic 30 is ready
for use, Faults must
be corrected by
_________________
Do not use
Basic 30 until the
faults are
corrected. Send
Basic 30 to Medela
Service center.