Metrix ScopiX IV OX 9062 Manual De Instrucciones página 93

Tabla de contenido

Publicidad

Programación
DISPlay[:WINDow]:CURSor
:STATe
DISPlay[:WINDow]:CURSor
:TIME{[1]|2|3}:POSition
DISPlay[:WINDow]:CURSor
:VOLT{[1]|2}:POSition
DISPlay[:WINDow]:TRACe
:FORMat
DISPlay[:WINDow]:TRACe
:MODE
DISPlay[:WINDow]
:TRACe:STATe{[1]|2|3|4}
DISPlay[:WINDow]
:TRACe:X[:SCALe]
:PDIVision
(Command/Query)
The DISP:CURS:STAT <1|0|ON|OFF> command activates or inhibits the manual
measurements.
1|ON: activates the manual measurements
0|OFF: inhibits the manual measurements
To the question DISP:CURS:STAT?, the instrument returns the activation status of the manual
measurements.
(Command/Query)
The DISP:CURS:TIME{[1]|2|3}:POS <position|MAX|MIN> command
sets the horizontal position of the selected manual cursor.
<position> is a value in format NRf, it may be followed or not by a multiple and the unit.
By default the value is expressed in second.
This command acts on the manual cursors represented on the screen by the X-symbol
accompanied by an index (1, 2 or ).
To the question DISP:CURS:TIME{[1]|2|3}:POS?, the instrument returns the horizontal
position of the selected manual cursor.
Response format:
<measured value><NL>
value in format <NR3> expressed in second.
(Query)
To the question DISP:CURS:VOLT{[1]|2}:POS?, the instrument returns the horizontal position
of the selected manual cursor.
This command acts on the manual cursors represented on the screen by the X-symbol
accompanied by an index (1, 2).
Response format:
<measured value><NL>
value in format <NR3> expressed in volt.
(Command/Query)
The DISP:TRAC:FORM <A|XY> command selects the display mode of the instrument.
A validates the Oscilloscope display mode : Y = f(t)
XY validates the XY display mode : Y = f(x)
To the question DISP:TRAC:FORM?, the instrument returns the active display mode.
(Command/Query)
The DISP:TRAC:MODE <NORMal|ENVelope> command selects the display mode.
NORMal validates the Vector display mode.
ENVelope validates the Envelope display mode.
To the question DISP:TRAC:MODE?, the instrument returns the active display mode.
(Command/Query)
The DISP:TRAC:STAT{[1]|2|3|4} <1|0|ON|OFF> command validates or devalidates the
selected signal.
To the question DISP:TRAC:STAT{[1]|2|3|4}?, the instrument returns the validation status of
the selected signal.
(Command/Query)
The DISP:TRAC:X:PDIV <scale|MAX|MIN|UP|DOWN> command sets the value of the time
base.
<scale> is a value in format <NRf> , it may be followed or not by a multiple and by the unit.
By default, the value is expressed in second.
Example: to get a time base of 1 µs, following values can be entered: 1E-3ms or 1E-6 or
0.000001s or 0.000001 or else 1us.
To the question DISP:TRAC:X:PDIV?, the instrument returns the value of the time base.
Response format:
<measured value><NL>
value in format <NR3> expressed in second.
93