Daily Inspection

Check before use every time:
1
Seal is undamaged.
Lid is closed and handle bolted.
No cracks, holes or other damage
more than 1.5 mm in depth can be
seen on the casing.
2
Indicator is dark blue in colour.
If there is a significant loss of blue
colour (50 % of the particles have
changed colour from dark blue to
light blue or neutral) the oxygen self-
rescuer must not be used any more.
Pack the oxygen self-rescuer and send to Dräger Safety.

Assembly of Shoulder Strap

Oxy K 30 HS/HS KD
To complete the training unit ready for use mount the shoulder strap
:
as shown
Loop shoulder strap in back plate
Loop short end of shoulder strap in
slot of back plate (1).
Loop shoulder strap in clamp buckle
Mount the ends of shoulder strap
into the clamp buckle and pull the
ends of shoulder straps as far as
possible (2).
Close clamp buckle (3).

Advice on Use

Do not open the oxygen self-rescuer until it is required for use.
Humidity in the atmosphere may penetrate the open oxygen self-rescuer
and impair its function.
Inspect the oxygen self-rescuer daily.
Use the oxygen self-rescuer only once.
To avoid risk of fire make sure that no flammable materials (petrol,
grease, solvents etc.) can get into the oxygen self-rescuer before or
during operation.
There is also danger of ignition if the oxygen-releasing chemicals come
into contact with combustible substances such as coal when the
oxygen self-rescuer is damaged.
Opened units are classified as used and it is not allowed to store them.
They have to be recycled (see "Disposal" on page 12).
For routine training purposes, use the training unit
Oxy K 30 HWT/HST
(see "Order List" on page 15).
1) not CE-approved.
2
1
1
2
3
1)

Donning Procedure

If the oxygen self-rescuer is not donned correctly, there could be a
delay in starting in an emergency
Perform the following donning steps according to the described
order.
Oxy K 30 HW
Opening the case
Take the yellow opener.
Move the handle downwards until lid
removes from oxygen self-rescuer.
Remove lid.
Putting on the unit
Take neck strap and pull out oxygen
self-rescuer with the second hand at
the housing carrying frame upwards
from holder.
Turn the oxygen self-rescuer in a way
that the housing carrying frame
shows to the chest.
Pull red rubber neck band over head
and hang around neck.
Pull the neck straps to position the
oxygen self-rescuer correctly.
Oxy K 30 HS/HS KD
Opening the case
Take the yellow opener.
Move the yellow opener away from
body until lid removes from oxygen
self-rescuer.
Remove the lid from oxygen
self-rescuer in straight direction
(straight away from body).
Remove lid.
Putting on the unit
Adjust shoulder strap around neck.
Pull both ends of shoulder strap to
the end.
.
11