SECTION 2—ASSEMBLY
Attaching the Battery Charger Mounting Bracket to the Wall
NOTE: For this procedure, refer to FIGURE 2.3.
NOTE: This procedure only applies to RPS350‐1E models.
1.
Place the battery charger mounting bracket on the wall
at the desired position.
2.
With a pencil, mark the middle hole position.
3.
Measure down 6½ inches from the pencil mark and
drill one mounting hole.
4.
Install the bottom mounting screw until there is an
approximate 1/8‐inch gap between the screw head and
the wall.
5.
Put the battery charger mounting bracket onto the bottom
mounting screw.
6.
Drill the other two mounting holes.
7.
Screw the mounting screws through the battery charger
mounting bracket and into the wall. Tighten securely.
CAUTION
Make sure there is an audible click when mount-
ing battery on the battery charger to confirm
proper mounting. Otherwise, injury or damage
may occur.
8.
Put the battery charger in place on the mounting bracket.
9.
Plug the battery charger into an electrical outlet.
NOTE: An LED will illuminate when power is applied to battery
charger.
Stand Assist Patient Lift
Mounting Bracket
Mounting
Screws
FIGURE 2.3 Attaching the Battery Charger Mounting Bracket
10
BOTTOM Mounting Screw
to the Wall
Part No. 1078985