CALIBRATING THE BOTTLE TEMPERATURE
NOTE: A digital thermometer is required for bottle temperature sensor calibration. Check that
the bottle temperature probe is disconnected from the bottle and so capable of reading the
ambient temperature.
Switch on the machine. From the CALIBRATION MENU:
BOTTLE PRESSURE
BOTTLE PRESSURE
BOTTLE TEPERATURE
Select BOTTLE TEMPERATURE and the following screen is displayed:
Check that the temperature shown on the display is that also read on the external
thermometer. If necessary, press START to modify the machine value:
Type in the temperature read on the digital thermometer and press ENTER to confirm.
Replace the temperature probe on the bottle.
INSTRUCTIONS MANUAL
BOTTLE TEMPERATURE
XX.X °C
START to change
STOP to exit
BOTTLE TEMPERATURE
XX.X °C
XXX.X °C
Press START
-33-