Emission test
Compliance
RF emission - CISPR
11.
10.2 - Magnetic and electromagnetic immunity
The C
2 is intended for use in the electromagnetic environment specified in the table below.
ELALUX
The user and/or installer must ensure that the C
Immunity test
Magnetic field at 50 Hz.
3 A/m
IEC61000-4-8
Electrostatic discharge
± 6 KV contact
(ESD)
± 8 KV air
IEC 61000-4-2.
Electrical fast transients
± 2 KV for power supply
IEC 61000-4-4.
lines
Surges
± 1 KV differential mode
IEC 61000-4-5.
± 2 KV common mode
<5% UT
0.5 cycles.
Voltage dips, short
40% U
interruptions and voltage
cycles
variations
70% U
IEC 61000-4-11.
25 cycles
<5% U
250 cycles
10.3 - Electromagnetic immunity / mobile radiofrequency equipment
The C
2 is intended for use in the electromagnetic environment specified in the table below.
ELALUX
The user and/or installer must ensure that the device is used in such an electromagnetic environment.
Immunity test
Portable and mobile radiofrequency communications devices must not be used near the C
Conducted disturbance,
radiofrequency fields.
IEC61000-4-6
Radiated radiofrequency
electromagnetic field.
IEC61000-4-3
The electromagnetic field strengths of fixed radiofrequency emitters, as determined by an electromagnetic environment measurement (a), must be
less than the compliance level in each frequency range (b).
Interference may occur near equipment marked with the symbol below:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
C
2 uses RF energy for internal operation.
ELALUX
Therefore, its radiofrequency emissions are very low and are not likely to cause any
interference in nearby equipment.
Group 1
Class B
C
2 is suitable for use in all establishments, including domestic establishments and
ELALUX
those directly connected to the public low voltage power supply network that supplies
buildings used for domestic purposes.
IEC 60601 test level
3 A/m
± 6 KV contact
± 8 KV air
± 2 KV for power supply
lines
± 1 KV differential mode
± 2 KV common mode
(>95% dip in U
) for
<5% U
T
T
0.5 cycles.
(60% dip in U
) for 5
40% U
T
T
cycles
(30% dip in U
) for
70% U
T
T
cycles
(>95% dip in U
) for
<5% U
T
T
250 cycles
IEC 60601 test level
recommended and calculated according to the frequency and power of the emitter.
3 V/m
150 KHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Electromagnetic environment - guidance
2 is used in such an electromagnetic environment.
ELALUX
Compliance level
The intensity of the magnetic field should be equivalent to that of
a typical commercial or hospital environment (hospital, clinic).
Floors must be wood, concrete, cement or tiled.
If floors are covered with synthetic material (carpet, etc.), the
relative humidity must be at least 30%.
Power quality should be that of a typical commercial or hospital
environment.
Power quality should be that of a typical commercial or hospital
environment.
(>95% dip in U
) for
T
T
Power quality should be that of a typical commercial or hospital
(60% dip in U
) for 5
T
T
environment.
If the use of the C
(30% dip in U
) for 25
T
T
power cut, it is recommended that the product be powered from
a separate power supply (UPS, etc.).
(>95% dip in U
) for
T
T
aCompliance level
Recommended separation distance:
3 V/m
d = 1.2
d = 1.2
d = 2.3
3 V/m
Where P is the maximum power rating of the emitter in watts (W)
according to the manufacturer's specifications and d is the
recommended minimum separation distance in meters (m).
9
Electromagnetic environment - guidance
2 requires continued operation during a
ELALUX
Electromagnetic environment - guidance
2 (including its cables) at a distance less than that
ELALUX
80MHz to 800MHz.
800MHz to 2.5GHz.