Safety Instructions - VOCO Celalux 2 Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 24
1 – SAFETY INSTRUCTIONS
Warning: Before using your C
read the whole of this Safety Instructions
section carefully. The manufacturer's warranty is
applicable only if these instructions concerning the
operation and the safety of the unit have been properly
applied. All these safety measures assume a good level
of knowledge of dental techniques and photo-
polymerization, and the more specific knowledge
regarding the use of the C
Operating Manual.
1) This unit must be used in strict accordance with the
instructions contained in this Operating Manual. The
manufacturer declines any liability if these rules are not
followed or if the unit is used for any other application.
2) Before connecting the unit, make sure that the power
voltage is compatible with the voltage marked on the
power adapter used to charge the curing light. A
different voltage would damage the unit and could
result in injury to the patient and/or the operator.
3) The light radiation produced by this type of unit can
be dangerous and must not be pointed at the eyes,
even if the practitioner or the patient is wearing
suitable protective eyewear. The light produced by
this equipment must be directed only at the zone to
be treated in the oral cavity.
4) Heat: As with any curing device, the emission of high
intensity light causes a temperature rise which can
damage the dentine or the soft tissue during
prolonged exposure. Therefore, the Celalux 2 should
be used only by experienced and qualified dental
professionals. Do not exceed an exposure time of 40
seconds in one location. Avoid direct contact with the
gingiva, the mucosa and the skin. Using a dam during
the polymerization should limit the temperature rise.
5) Any condensation inside electrical equipment can be a
hazard. If the curing light has to be moved from a cool
place to a warm place, do not use it immediately: let
it reach the new room temperature first.
6) Do not replace the battery during use.
7) Comply with all the instructions given in this Manual.
Do not insert or attempt to insert metal objects into
the unit or its base station, as this could cause an
2, please
ELALUX
2 contained in this
ELALUX
electrical discharge.
8) The manufacturer declines all liability if damaged
parts or accessories are not replaced exclusively by
the manufacturer's products. In particular, the use of
other light guides, power adapters or batteries could
be dangerous for the patient and the operator.
9) In the case of a fault, unplug the base station
immediately from the power grid and make sure that
no one can use the curing light before it is checked
by the manufacturer or dealer. The fault may be the
result of non-compliance with the safety rules or of a
technical problem in the unit.
10) Do not use the unit near a heat source. Do not use
solvents, detergents or flammable products, which
may damage the unit or cause short-circuits.
11) Only the unit manufacturer or dealer is qualified to
carry out repairs.
12) All lights used for photo-polymerization, including
the C
2, must not be used on persons who are
ELALUX
suffering or have suffered from photo-biological
reactions
(including
erythropoeitic protoporphyria) or who are receiving
treatment including photosensitizing medications
(including methoxsalens or chlorotetracycline).
13) Persons (practitioners or patients) who have
previously suffered from a retina or crystalline lens
condition or who have undergone eye surgery, in
particular
for
ophthalmologist before using the C
the eye specialist agrees, prudence is strongly
recommended, because the intensity of the light
could cause accidents. Class 2M protective eyewear,
suitable for use with units emitting radiation at
wavelengths between 420 and 480 nm, should be
worn at all times.
14) C
2 must not be used if the patient and/or the
ELALUX
operator has a cardiac stimulator or any other active
implant (e.g. a cochlear implant).
15) Any variation of the power supply voltage or the
magnetic field outside the limits in force could
switch the curing light to automatic operation or
interfere with its operation.
16) Electromagnetic interference: the unit complies with
the applicable standards (IEC 60 601-1-2) for emission
(radiated electrical field and interference with the
supply voltage) and for immunity (protection against
4
urticaria
solaris
cataracts,
must
see
ELALUX
or
their
2. Even if

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido