Starting, fi nal test and users instruction
Install and connect the aspirator, select the position ON on
the switch, which will light up after one of the dental units has
started working.
At this point aspiration will start.
To check if Turbo-Smart is working correctly, it is advisable to carry
out the dynamic tests (see fi g. D, page 93 and the Turbo-Smart
working diagram fi g. E, page 94).
Users must be instructed on the aspirator use and routine
maintenance of new, not used, and therefore not yet contami-
nated, machines.
Show users how to follow the Turbo-Smart working phases
on the display, to interpret danger warnings and to carry out
routine maintenance using Puli-Jet plus (A) - (by means of
Pulse-Cleaner) (B) - and Antifoaming Tablets (C).
Working
The aspirating unit (9) (through pipe 10) creates vacuum inside
the centrifugal separator (11).
The fl uid coming from the dental units enters the centrifugal
separator (11) from the collecting pipe (2).
The centrifugal separator separates air from liquids: air is
exhausted outside through the pipe (4) whereas liquids (in the
version without amalgam separator) are drained to the sewage
through the waste pipe connected to the tube-holder (6).
The centrifugal separator (11) is starting before the aspirating
unit (9), this allows to drain the liquids which might collect inside
the centrifugal separator before aspiration starts. Moreover,
when the machine is switched off a timer (adjustable) is keep-
ing the motor running for min.10'' - max. 120''.
Amalgam Separator ISO 18
On request, Turbo-Smart can be supplied with the amalgam
separator "Hydrocyclone ISO 18" having its separate manual.