Alarms description

Alarm code
AC80
AC80-07
00
0
I14
2
I15
3
I00
32
I01
33
I02
34
I03
35
I04
36
I05
37
I07
39
S08
40
S09
41
S10
42
S11
43
I16
48

Important notices

• Appliances are guaranteed for one year from the date of sale, provided that the warranty slip is
returned to the manufacturer with date of sale, retailer's and customer's name.
• Warranty and manufacturer's liability cease in case appliances are treated with products which
are unsuitable or different from those recommended by the manufacturer and also in case appli-
ances are improperly used or tampered with operations of any kind carried out by people who
are not authorized by the manufacturer.
• The manufacturer, concessionaires, agents and authorized technicians are at customers' disposal
for advice and assistance and to supply literature, spare parts and anything useful.
• The manufacturer reserves the right to modify the products for improvements, for technical, nor-
mative and functional reasons or for problems due to the availability of products or semi-fi nished
products, without prior notice.
• Our updated manuals are available on the web site www.cattani.it.
We recommend they are consulted especially for updates about safety.
HOME
DESCRIPTION
Microcontroller memory alarm
Amalgam level > 95%
Amalgam level > 100%
Microcontroller memory alarm
Short-circuit due to one of the two
motors
Short-circuit before the motors' start-
ing
Condensers have not been charged Contact the technician (replace the
Overtemperature (temperature has
exceeded the max. allowed limit)
Blower overcurrent (the blower has
exceeded the current limit)
Condensers overvoltage (the con-
densers have exceeded the max.
voltage limit)
Centrifuge short-circuit
PC-Board short-circuit on the centri-
fuge output
Centrifuge instant overcurrent (the
centrifuge has exceeded the current
limit)
Centrifuge time delayed overcurrent
(the centrifuge has exceeded the
current limit)
The vacuum detecting pipe is discon-
nected
Contact the technician
Replace the amalgam container as
soon as possible
Replace the amalgam container
Contact the technician
Identify where the short-circuit is
coming from and eliminate it
Contact the technician (the PC-Board
is likely to be damaged)
PC-Board)
Ventilate the plant room
Check the operation of the blower
(seized or excessive friction)
Check the mains voltage (max. 260V)
Eliminate the short-circuit
Replace the PC-Board
Eliminate siphons in the piping or
check the centrifuge operation (seized
or excessive friction)
Eliminate siphons in the piping or
check the centrifuge operation (seized
or excessive friction)
Connect the vacuum detecting pipe
to the centrifuge
34
SOLUTION
INDEX

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Turbo-smart b