System Parameters Setup menu
To access this menu and/or make any changes, you are
required to enter the access password 0000456000 (see
instructions on page 30).
Vacuum set point
(Max.setting: 200 mbar)
To change any parameters in this menu:
Scroll with
through to display the required parameter.
Press the
I
Enter Key to enable the cursor. The figure can
be changed using the arrow keys
.
Once the figure has been set to the desired figure, press the
Enter Key again to confirm and continue.
Blower Current Limit
(Max. setting for A: 5,5 A
Max. setting for B: 7,5 A)
Turbo-Smart Off Delay Time
(Max. setting: 300 sec.
Min. setting: 10 sec.)
Pump (Centrifuge) Off Delay Time
(Max. setting: 30 sec.
Min. setting: 5 sec.)
Control Panel Fan Off Delay Time
(Default setting: 300 sec.
Max. setting 1800 sec.)
Running Options
This enables or disables the amalgam level sensor.
Set to 0: without amalgam separator
Set to 1: with amalgam separator
Drive Commands