Caution: Do not depress the button during use, the handle will release from forceps.
5. Prior to handle removal, ensure that the biopsy sample is retrieved.
6. At the completion of the procedure, detach the handle according to instructions as
follows:
a) With the button depressed, retract the finger spool over the button.
b) Secure handle in this position by releasing the button (Fig. 2). The forceps shaft may
now be removed from the handle.
7. Remove BIGopsy® forceps from endoscope.
8. Upon completion of the procedure, dispose of the device per institutional guidelines for
biohazardous medical waste.