b. Speed adjustment
SPEED +/-: Sets motor rotation speed (value in rpm x 20 steps). Speed values depend on contra-angle reduction
rate. See the following table:
Avoid excessive speed. Set the speed to a safe maximum for the application.
c. Torque adjustment
TORQUE +/-: increases or decreases the torque values (24 steps). Torque values are in Ncm and depend on
contra-angle reduction rate. See the following table with the handpiece 20:1:
3.0
21.0
39.0
57.0
Torque accuracy is guarantee only with a contrangle correctly maintained. The torque
values are only for handpieces that operate properly. Micro motor stops when set torque is
reached. The torque value in counterclockwise rotation is always = Max Ncm
d. Flow
FLOW: Selects variable irrigation pump flow rate. See the following table
0 = no irrigation
e. Reverse
REVERSE: Switch between forward and reverse motor rotation (when in reverse, console will emit a beeping
tone).
f. Motor
MOTOR ON/OFF: Switches motor ON and OFF (if enabled).
g. Pump
PUMP ON/OFF: Switches pump ON and OFF (if enabled).
Always check the values before use. Danger of injury for incorrect use.
h. Footswitch
The most important functions can be activated either by the control panel and the variable-speed footswitch.
(Fig. 4).
Reduction Rate
1:1
16:1
20:1
24:1
32:1
64:1
80:1
6.0
9.0
24.0
27.0
42.0
45.0
60.0
63.0
1 = MIN
Max. RPM
40,000
2500
2000
1666
1250
625
500
Ncm
12.0
30.0
48.0
66.0
2 = MED
Min. RPM
400
25
20
16
12
6
5
15.0
18.0
33.0
36.0
51.0
54.0
69.0
MAX
3 = MAX
14