Burst/Fast transient
2kV for power supply
EN 61000-4-4
lines
Surge
1kV differential mode
EN 61000-4-5
2kV common mode
Voltage dips, short
5%Ut
interruptions and
(95% dip in U
voltage variation on
40% Ut
power supply input
(60% dip in U
lines.
70% U
(30% dip in U
EN 61100-4-11
5% U
(95% dip in U
Power frequency
(50/60 Hz)
magnetic field
3A/m
EN 61100-4-8
2kV for power supply
lines
1kV differential mode
2kV common mode
5%Ut
) for 0,5 cycle
(95% dip in U
t
40% Ut
) for 5 cycle
(60% dip in U
v
70% U
t
) per 25 cycle
(30% dip in U
t
5% U
t
) for 5 sec
(95% dip in U
t
3A/m
Quality of the Main power source should be that
of a typical commercial or hospital environment
Quality of the Main power source should be that
of a typical commercial or hospital environment
Quality of the Main power source should be that
of a typical commercial or hospital environment If
) for 0,5 cycle
t
the user of the I-SURGE requires continued
operation during power main interruptions, it is
) for 5 cycle
recommended than the I-SURGE be powered from
v
an uninterruptible power supply or battery.
t
) per 25 cycle
t
t
) for 5 sec
t
Power frequency magnetic fields should be at a
typical commercial or hospital establishment
level.