Satelec Acteon I-Surge Manual De Uso página 20

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 56
•Wrong voltage.
•Broken connections inside.
• Transformer damaged.
• PCB faulty.
• Filter supply defective.
• Power interrupted.
Fault detected: The central unit turns off and stops working suddenly
Possible cause
• Power cable damaged.
• Filter damaged.
• Transformer damaged.
• PCB damaged.
• Connections inside interrupted.
Fault detected: The central unit doesn't switch off
Possible cause
•Power filter damaged.
Fault detected: The foot pedal doesn't start the motor
Possible cause
• Micro Motor cable not connected properly Check to see that the Micro Motor cable is securely plugged into the connector on
• Footswitch cable is not connected properly Check that the footswitch cord is plugged into the connector on the front left of
•Male connector of the Micro motor
damaged.
•Pins of the connector damaged
•Wrong voltage.
•Foot pedal damaged.
Fault detected: It is not possible to change "Program", FWD/REV and Flow by using the footswitch.
Possible cause
• Footswitch damaged.
•Metal pins damaged in the connector
Fault detected: The display is not working properly
Possible cause
•Contra-angle damaged.
•Wrong voltage.
Verify that the voltage matches the power supply. If it does, try to switch the
central unit off and then on again, or connect the device to another working
electrical outlet. If the unit does not turn on, verify the integrity of the fuses in
the unit.
Immediately discontinue use of the device and contact SATELEC after-sales
service.
Solution
After checking the correct connection to the electrical outlet and the absence of
liquid in the unit, try to power the central unit. If the device does not turn on,
immediately discontinue use of the device and contact SATELEC after-sales
service.
Solution
Unplug the power cord from the electrical outlet and the central unit. Then
contact SATELEC after-sales service.
Solution
the front right of the central unit, disconnect and reconnect.
the central unit, disconnect and reconnect.
Unplug the Micro Motor and make sure that the male connector is not damaged,
wet or damp.
Unplug the Micro Motor and the footswitch cable, make sure the metal pins on the
male connector are intact and not damaged and have not broken off and/or
remained stuck in the female connector mounted on the unit.
Verify that the voltage matches the power supply. If it does, try to switch the
central unit off and then on again, or connect the device to another electrical
outlet.
Check the ability to start the Micro motor from the central unit keyboard, pressing
the [Motor ON/OFF switch]. If so, contact SATELEC after-sales service.
Solution
Check the possibility of changing the parameters by using the unit keyboard.
Contact SATELEC after-sales service.
Solution
Check the integrity of the Contra-angle.
Verify that the voltage matches the power supply. If it does, try to switch the
central unit off and then on again, or connect the device to another electrical
outlet.
20

Publicidad

Tabla de contenido
loading

Tabla de contenido