3 Operating
K
Operation model/program selction
3
With this key you can select the eight preset modes
. According to mode,
the keys A, B, E, and G are activated or deactivated (see table " Aavailable
functions - overall view"). With each depression of the key (K), you switch to
the next mode/program.
After switching on the device, it is in the factory setting, the mode CONST and
shows the following values: 2 Hz, 50 µs, 1. bi-phasic current selection.
CONST
In this mode you can set all values for frequency (E), pulse width (G) and
current selection (B) by yourself and constantly keep them.
BURST
This mode generates pulse packets of the preset frequency that are switches
on and off twice a second. The pulse width is adjustable, you can select the
current selection as well.
MODUL I
In this mode, the frequency of the set pulses (pulse width and current selection
selectable) is step by step increased from the lowest frequency (2 Hz) to the
highest (200 Hz) frequency. The frequency is increased every 10 seconds by one
step. After 90 seconds, the device starts again with the lowest frequency.
MODUL II
In this mode, the pulse width is changed from the highest to the lowest value
and back. This process is repeated every six seconds.
With the frequency setting E you can set the frequency of the pulse packets.
PROG A
The intensity value for the current set by you is lowered to 70% and raised in
form of a ramp within 10 seconds. This raising and lowering is not visible in the
display. You only feel it as rising and lowering tingling. The settings for current
selection, frequency and pulse width are selectable and are not influenced.
31