Electromagnetic Compatibility - Stryker iSwitch Manual De Operación Y Mantenimiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 58

Electromagnetic Compatibility

Like other electrical medical equipment, the iSwitch requires special precautions
to ensure electromagnetic compatibility with other electrical medical devices. To
ensure electromagnetic compatibility (EMC), the iSwitch must be installed and
operated according to the EMC information provided in this manual.
Note
The iSwitch has been designed and tested to comply with IEC
60601-1-2:2001 requirements for EMC with other devices.
Caution
The iSwitch may be interfered with by other equipment,
including portable and mobile RF communication equipment,
even if such equipment meets the applicable emissions
requirements.
WARNING Do not use cables or accessories other than those provided
WARNING If the iSwitch is used adjacent to or stacked with other
Guidance and Manufacturer's Declaration: Electromagnetic Emissions
iSwitch is intended for use in the electromagnetic environment specified below. The customer or the user of
Emissions test
RF emissions
CISPR11
RF emissions
CISPR11
Harmonic emissions
IEC61000-3-2
Voltage Fluctuations/
flicker emissions
IEC61000-3-3
17
with the iSwitch, as this may result in increased
electromagnetic emissions or decreased immunity to such
emissions.
equipment, observe and verify normal operation of the
iSwitch in the configuration in which it will be used prior to
using it in a surgical procedure. Consult the tables below for
guidance in placing the iSwitch.
iSwitch should ensure that it is used in such an environment.
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic Environment - guidance
The iSwitch must emit electromagnetic energy in
order to perform its intended function. NEarby elec-
tronic equipment may be affected.
iSwitch is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.