Guidance and Manufacturer's Declaration: Electromagnetic Immunity
iSwitch is intended for use in the electromagnetic environment specified below. The customer or
the user of iSwitch should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Conducted RF
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast, cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the iSwitch system is used
exceeds the applicable RF compliance level above, the iSwitch system should be observed to verify normal opera-
tion. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocat-
(b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
19
3 Vrms
3 V/m
80MHz to 2.5 GHz
and reflection from structures, objects, and people.
ing the iSwitch unit.
Compliance
Electromagnetic Environment: Guidance
Level
Portable and mobile RF communications
equipment should be used no closer to any
part of the iSwitch system, including its cables,
than the recommended separation distance
calculated from the equation applicable to the
Recommended Separation Distance
3 V
3 V/m
d
d
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the rec-
ommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
(a)
, should be less than the compliance level in
Interference may occur in the vicinity of
equipment marked with the following symbol:
frequency of the transmitter.
d
=
1.17 P
=
1.17 P
80 MHz to 800 MHz
800 MHz to 2.5 GHz
=
2.33 P
(b)
each frequency range
.