CLEANING AFTER EACH USE
Following the cleaning instructions after each use will prevent any
remaining medication in the bottle from drying, adhering to the mesh
cap, and resulting in the device not nebulizing effectively.
Wash the Medication Bottle, Mesh Cap, Mask, Mouthpiece, and Mask
Adapter after each use.
1. Remove the Mouthpiece or Mask and the Mask Adapter from the Main Unit.
2. Remove the Medication Bottle from the Main Unit.
3. Open the Medication Bottle and discard any remaining medication.
4. Attach the Medication Bottle to the Main Unit. Open the Mesh Cap.
5. Pour a small amount of distilled water into the Medication Bottle
and close the Mesh Cap.
6. Turn on the device to nebulize the distilled water for 1 to 2 minutes to
remove residual medication from the mesh holes.
7. Turn off the device and remove the Medication Bottle from the Main Unit.
8. Remove the Mesh Cap from the Medication Bottle and discard any
remaining distilled water from the Medication bottle.
9. Rinse the Medication Bottle, Mesh Cap, Mask, Mouthpiece, and Mask
Adapter with distilled water.
Caution
To prevent damage to the device:
• Do not rinse or immerse the Main Unit in any liquid.
• Do not wash or rinse any of the parts under strong running water.
• Do not touch the mesh with your hand or any other object.
10. Gently wipe off excess water with a soft clean cloth or allow the parts
to air dry in a clean environment.
11. Assemble the device. Store the device in the Storage Case
or in a clean environment.
14