Daily Disinfecting - Omron MICRO Air NE-U22V Manual De Instrucciones

Vibrating mesh nebulizer
Ocultar thumbs Ver también para MICRO Air NE-U22V:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Disinfect the Medication Bottle, Mesh Cap, Mask, Mouthpiece, and Mask
Adapter after the last treatment of each day.
1. Make a solution of 1 part white vinegar and 3 parts distilled water.
Caution
Do not use household bleach. The mesh will rust.
2. Lift open the Mesh Cap and pour a small amount of disinfecting
solution into the Medication Bottle.
3. Turn on the device to nebulize the disinfecting solution for
1 to 2 minutes.
4. Turn off the device and remove the Medication Bottle from
the Main Unit.
5. Remove the Mesh Cap from the Medication Bottle and discard any
remaining disinfecting solution from the Medication Bottle.
6. Soak the Medication Bottle, Mesh Cap, Mask, Mouthpiece, and Mask
Adapter in the disinfecting solution for 10 to 15 minutes.
7. Rinse the Medication Bottle, Mesh Cap, Mask, Mouthpiece,
and Mask Adapter in distilled water.
Caution
To prevent damage to the device:
• Do not rinse or immerse the Main Unit in any liquid.
• Do not wash or rinse any of the parts under strong running water.
• Do not touch the mesh with your hand or any other object.
8. Gently wipe off excess water with a soft clean cloth or allow the parts
to air dry in a clean environment.
9. Assemble the device. Store the device in the Storage Case
or in a clean environment.

DAILY DISINFECTING

15

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Micro air ne-u22vac

Tabla de contenido